HOME >FAQ

  • About properties(14)
  1. Please tell me how to convert precision display into ± kPa.

Precision display %FS is conversion value when you regard output voltage as 100% at rated pressure.
Therefore, the conversion into kPa is made by “Rated pressure x %FS”.

For example, when you want to convert rated pressure 98.1 kPa ± 0.3%FS,the calculation is ±98.1 kPa x 0.003 = ±0.2943 kPa.